Interior light - Burned brightly when switch was turned on.
Shaker motor - Ran at various speeds in accordance with adjustment of speed knob.
Heating Coils - turned on and provided heat when switch and thermostat were turned to heating.
Cooling Pump - turned on and cooled the unit when switch and thermostat were turned to cooling.
It should be noted that the main fuse that came with this unit was the wrong amperage. As a result of this, we blew several fuses before ascertaining that a minimum of ten amps was required to run the heating and cooling circuit. If you know the rating required for the main fuse in this unit please let us know so we can test this unit with the correct fuse.
Furthermore, the compressor for the cooling circuit is sitting loose under the unit instead of being mounted to the frame.