The API 200A measures the concentration of NO, NOx, and NO2 by measuring the light intensity of the chemiluminescent reaction of nitric oxide and ozone. This analyzer is microprocessor controlled, and continually checks operating parameters such as temperature, flow, and critical voltages at the PMT and converters. Measurements are automatically compensated for temperature, pressure and flow rate. Stability is enhanced by an auto-zero circuit that continuously provides a true zero reference. It features a built-in data acquisition system, an RS232 port to allow control, programming, or data upload to a PC, and built-in test ,diagnostic, and calibration modes that allow the user to check and maintain optimum performance.
In 1ppb increments from 50ppb to 20,000ppb
Single, Independent, AutoRange
1 ppb/7 days Span Drift <0.5% FS/7 days Lag Time 20 sec
500 cc/min. 10% Linearity 1% of full scale
HxWxD 7" x 17" x 23.6" (18cm x 43cm x 61cm)
43 lbs (20 kg) w/external pump
55 lbs (25 kg) w/internal pump
100 V ~ 50/60 Hz, 120V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 125 watts
Installation Category (Over-voltage Category) II Pollution Degree 2
1 part in 1024 of selected voltage or current range
12 Status Outputs from opto-isolator